Measuring point multiplicator Type TMU104V
Measuring Point Multiplicator, 1 input for sensors Pt 100, Pt1000, KTY 83/84, thermocouples, 4 insulated outputs Pt100, via interface RS485 (Modbus RTU) fourfold Pt100 simulator
Product description
The measuring point multiplicator TMU104V measures the temperature at a connected sensor and transduces it into 4 insulated signals Pt100 (RTD). Via interface RS485 it can be used as a simulator for up to 4 signals Pt100.
Overview of functions
- Measuring input Pt100 (RTD), Pt1000, KTY83 / 84 in 2- or 3-wire
- Measuring input thermocouple (types B, E, J, K, L, N, R, S, T)
- Measuring range -199…+850 °C
- 4 insulated outputs signal Pt 100 (resistance- signal), connection in 2-, 3- or 4-wire
- Digital display, 3 digits, resolution 1 °C (-19.9 … 99.9 °C: resolution 0,1 °C)
- Storing of MIN- and MAX- values
- Universal supply voltage AC/DC 24-240 V
- Interface RS485 (protocols ZIEHL and Modbus RTU)
- Housing for DIN-rail or wall-mount, 105 mm wide, mounting height 55 mm
Measuring Point Multiplicator and Transducer
The temperature of the sensor (resistance or thermocouple), connected to the input, is available as signal Pt100 (RTD) at 4 insulated outputs.
Thus allows the connection of other sensors than Pt100 to inputs for Pt100 at other devices.
Normally only one input can be connected to a temperature sensor. With help of TMU104 up to 4 devices (controls, displays, monitoring devices) with inputs Pt100 can be connected to one sensor at the same time.
Simulator for Pt100
Controlled via interface RS485 (protocol Modbus RTU) the TMU1004V can simulate up to 4 sensors Pt 00 (RTD). This allows the application in equipment, that makes automatic tests and calibrations at devices and installations with several inputs Pt100.
Technical data
Rated supply voltage Us:
AC/DC 24 – 240 V
Tolerance
DC 20,4 - 297 V AC 20 - 264 V 50/60 Hz
Power consumption
< 2,5 W
Sensor input 1T1/1T2/1T3
Pt100, Pt1000 according to EN 60751:
Tolerance
±0,2 % of measured value ±0,5 K (KTY ±5 K)
Sensor current
≤0,6 mA
Temperature drift
< 0,04°C/K
Measuring time 2 wire connection/3 wire connection
≤ 330 ms/ ≤ 440 ms
Thermocouples according to EN 60 584, DIN 43 710
Temperature drift
< 0,01 %/K
Measuring error of the sensor wire
+ 0,25 μV / Ω
Reference junction
± 5 °C (after warmup-time of 15 minutes)
Measuring time
≤ 440 ms
Sensor output OUT 1 … OUT 4
Type
Pt100 according to EN 60751
Reaction time to sensor input
< 10 ms
range of sensor current
200 μA … 5 mA
Type of connection
2-, 3-, 4-Leiter-Schaltung
Tolerance
< 200 Ω = ±0,2 % of simulated value
> 200 Ω = ±0,5 % of simulated value
Test conditions (EN 61010-1)
Rated impulse voltage
4000 V
Overvoltage category
III
Contamination level
2
Rated insulation voltage Ui
Ui 300 V
On-time
100%
Galvanic insulation / Test-voltage
Us – OUT1…4, Input, RS485 DC 3820 V
OUT1…4 – Output, RS485 DC 1000 V
OUT1 – OUT2 – OUT3 – OUT4 DC 1000 V
No galvanic insulation
Input – RS485
EMC-tests
EN 61326-1
Burst
EN 61000-4-4 +/-4 kV
Pulse 5/50 ns, f = 5 kHz, t = 15 ms, T = 300 ms
SURGE
IEC 61000-4-5 +/-1 kV
Impulse 1,2/50 μs (8/20 μs)
discharge of static electricity
IEC 61000-4-2 +/-4 kV contact, +/- 8kV air
Environmental conditions
Ambient temperature range
-20 °C ... +65 °C
Storage temperature range
-20 °C …+70 °C
Altitude
up to 2000 m
Climatic conditions
5-85% rel. humidity, no condensation
External wiring temperature range
-5 °C …+70 °C
Vibration resistance EN 60068-2-6
2…25 Hz ±1,6 mm
Housing
Housing
Design V6, switchgear mounting
Mounting height
55 mm
Width
6 TE
Dimensions (width x height x depth) (B x H x T)
105 x 90 x 58 mm
Wire connection, one wire
each 1 x 0,14 mm2 – 2,5 mm2
Stranded wire with insulated ferrules
each 1 x 0,14 mm2 – 1,5 mm2
Strip length
min. 8 mm
Torque
0,5 Nm (3,6 lb.in)
Protection class housing
IP 30
Protection class terminal
IP 20
Installation
Snap mounting on mounting rail 35 mm according to
EN 60 715 or with screws M 4
Weight
approx. 200 g